Medical Devices Regulations
SOR/98-282In forceCurrent to September 30, 2026 (Justice Canada: October 19)15 amendments not yet in forceHistoryRegistered on May 7, 1998Made under: Food and Drugs Act
Part 2 — Custom-Made Devices and Medical Devices to Be Imported or Sold for Special Access
Application
General
70 No person shall import or sell a Class III or IV custom-made device or a medical device for special access unless the Minister has issued an authorization for its sale or importation.
Authorization
71 (1) If a health care professional wishes to obtain a medical device referred to in section 70, the professional shall apply to the Minister for an authorization that would permit the manufacturer or importer of the device to sell, or to import and sell, the device to that professional.
(2) The application shall contain the following:
(a) the name of the device, its class and its identifier, including the identifier of any medical device that is part of a system, test kit, medical device group, medical device family or medical device group family;
(b) the number of units required;
(c) the name and address of the manufacturer or importer;
(d) the name, title and telephone number of the representative of the manufacturer or importer to contact for any information concerning the device;
(e) the diagnosis, treatment or prevention for which the device is required;
(f) a statement that sets out
(g) the name and address of each health care facility at which the device is to be used by that professional;
(h) the known safety and effectiveness information in respect of the device;
(i) a written undertaking by the health care professional that the professional will inform the patient for whom the device is intended of the risks and benefits associated with its use;
(j) the directions for use, unless directions are not required for the device to be used safely and effectively; and
(k) in the case of a custom-made device, a copy of the health care professional’s written direction to the manufacturer giving the design characteristics of the device.
- SOR/2023-19, s. 8
72 (1) The Minister shall issue an authorization referred to in subsection 71(1) to a manufacturer or importer if the Minister determines that
(a) the benefits that may be obtained by the patient through the use of the device outweigh the risks associated with its use;
(b) the health or safety of patients, users or other persons will not be unduly affected;
(c) a licensed device that would adequately meet the requirements of the patient is not available in Canada;
(c.1) a medical device for which the manufacturer of that device holds an authorization issued under section 68.12 and that would adequately meet the requirements of the patient is not available for sale in Canada; and
(d) the authorization is not being used by the manufacturer or importer to circumvent the requirements of Part 1 or 1.1.
(2) The authorization issued under subsection (1) shall specify
- SOR/2023-19, s. 9
Additional Information
73 If the information and documents submitted in respect of an application made pursuant to section 71 are insufficient to enable the Minister to determine whether the conditions set out in subsection 72(1) have been met, the manufacturer, importer or health care professional shall, at the request of the Minister, submit any further information relevant to the application that the Minister may request.
74 The Minister may, in respect of an authorization that has been issued,
(a) request the manufacturer, importer or health care professional to submit information in respect of the device if the Minister believes on reasonable grounds, after reviewing a report or information brought to the Minister’s attention, that the device for which the authorization has been issued no longer meets the conditions set out in subsection 72(1); and
(b) issue a written cancellation of the authorization, giving reasons, if
Labelling
Distribution Records
Incident Reporting
77 The health care professional referred to in subsection 71(1) shall, within 72 hours after becoming aware of an incident that involves the medical device for which an authorization has been issued under section 72 and that meets the following conditions, report the incident to the Minister and to the manufacturer or importer of the device and specify the nature of the incident and the circumstances surrounding it:
(a) the incident is related to a failure of the device or a deterioration in its effectiveness or any inadequacy in its labelling or in its directions for use; and
(b) the incident has led to the death or a serious deterioration in the state of health of a patient, user or other person or could do so were the incident to recur.
- SOR/2020-262, s. 19