7 A measuring apparatus that is required for measurement of electricity or gas or examination of meters shall be calibrated in the following manner:
(a) by testing its accuracy at points throughout the measuring range of the apparatus; and
(b) by tracing its accuracy at one or more points tested in accordance with paragraph (a) to reference standards kept by the National Research Council of Canada.